
70
Reference Manual
00809-0100-4811, Rev DA
Section 4: Basic Configuration/Start-Up
January 2015
Basic Configuration/Start-Up
4.4.2 Measurement units
Set transmitter units for level and temperature.
4.4.3 Reference gauge height
The Reference Gauge Height is the distance from the Upper Reference Point to the bottom of
the tank (see Figure 4-1 on page 62). When setting the Reference Gauge Height, keep in mind
that this value is used for all level measurements performed by the Rosemount 3300 Series
Transmitter.
The Reference Gauge Height must be set in linear (level) units, such as feet or meters, regardless
of primary variable assignment.
4.4.4 Probe length
The probe length is the distance from the Upper Reference Point to the end of the probe, see
Figure 4-1 on page 62. If the probe is anchored to a weight, do not include the height of the
weight. This parameter is pre-configured at factory. The probe length needs to be changed if,
for example, the probe is shortened.
4.4.5 Probe type
The transmitter automatically makes an initial calibration based on the type of probe that is
used. This parameter is pre-configured at factory and only needs to be set if the probe is
changed to another type. Select one of the following options:
Rigid Twin
Flexible Twin
Coaxial
Rigid Single 0.3 in. (8 mm)
Flexible Single
Coaxial HTHP
Coaxial HP/C
Rigid Single HTHP 0.3 in. (8 mm)
Flexible Single HTHP
Rigid Single PTFE
Flexible Single PTFE
Rigid Single HP/C 0.3 in. (8 mm)
Flexible Single HP/C 0.3 in. (8 mm)
HART Comm
1, 3, 1
HART Comm
1, 3, 2, 1
HART Comm
1, 3, 2, 2
HART Comm
1, 3, 2, 3
Commentaires sur ces manuels