Reference Manual
00809-0100-4811, Rev DA
Section 3: Installation
January 2015
Installation
3.5.7 Adjusting the probe length
1. Determine L, the desired probe length.
2. Determine n, the number of middle segments needed for the desired probe length. See
Tab le 3 -7 and Table 3-8 on page 46.
3. Calculate Y, the length of the bottom segment. See Tab le 3 -7 and Table 3-8 on page 46.
4. Continue as follows:
Length of bottom segment (Y) Action
Continue with step (7).
Do not use the bottom segment.
Continue with step (5) and cut the
bottom segment.
1. Add one extra middle segment to the
calculated n.
2. Continue with step (7).
L, desired probe length:
L
n, number of middle segments:
n
Y, length of bottom segment:
Y
Y < 0.4 in. (10 mm)
Y 0.4 in. (10 mm)
Y = 31.5 in. (800 mm)